Fluoride ISE sensors measure F- ion in solution the same way that a pH sensor measures hydrogen ions. A lanthanam fluoride crystal on the tip of the sensor develops a voltage that is proportional to fluoride ion activity. A reference electrode, which is part of the fluoride sensor, provides the other half of the sensing system, with the measurement made at a differential input amplifier. Since the activity of fluoride ions in solution are a function of pH and ionic strength, a small amount of buffer solution is added to the measured sample. This creates a stable condition in which the concentration of fluoride ion and the activity of fluoride ion are directly proportional.
In operation, a small amount of sample is pumped into the system and mixed with the buffer solution. The treated sample then flows to a chamber where the combination fluoride ISE and reference electrode is mounted. The fluoride ion concentration is measured in this chamber with changes in fluoride immediately reflected at the monitor. The treated sample is pumped back out of the measuring cell and into the drain side of the inlet overflow chamber. Fluoride monitors require very little maintenance. One gallon of buffer will operate the system for 25 days and sensor fill solution level should be checked weekly.
The Q46F provides continuous water quality monitoring of free fluoride concentration in potable water. The system employs a fluoride sensitive ion selective electrode (ISE) which provides reliable measurement down to 0.1 PPM and as high as 1000 PPM. A chemistry module provides sample conditioning for the sensor and the measured fluoride concentration is displayed on a separate electronics module that also provides alarm and analog output functions.
Features
Two isolated 4-20 mA outputs are standard. Default setting provides analog outputs for fluoride and temperature.
Standard PID control function assignable to one analog output.
Communication options for Profibus-DP, Modbus-RTU, or Ethernet-IP.
A sensor preamplifier module in the Auto-Chem enclosure allows separation between the monitor and chemistry unit by up to 300 feet.
Three SPDT relays are standard, with relay functions programmable for alarm, control, or trouble indication.
NEMA 4X (IP-66) enclosure is suitable for wall, pipe or panel mounting.
Back-lit large LCD display provides clear visibility in any lighting condition. A scrolling second line on the display provides additional information and programming prompts.
Specifications
Water & Wastewater Process Monitors and Sensors |
---|
Have any questions? Need on-site technical support with a product we supply? One of our technical sales representatives will be happy to help. Contact us today.
Please note that our new office address is 6470 Viscount Road, Mississauga, ON L4V 1H3.
Please note that our new office address is 6470 Viscount Road, Mississauga, ON L4V 1H3
Please note that our office address will change to 6470 Viscount Road, Mississauga, ON L4V 1H3 from November 14, 2023.